The 900 MHz XSC Wireless RTD Transmitter & External RTD Probe is ideal for applications where the internal temperature of an item needs to be monitored.
The 900MHz XSC Wireless Sensors are battery operated digital sensors with a microprocessor controlled 900 MHz FCC certified radio transmitter. The sensors have an on board time of day clock that allows them to spend most of the time in a low power quiescent state. At predetermined time intervals the clock will wake up the onboard microprocessor to transmit readings. This information is combined with a CRC-16 error check and transmitted in a very short data packet that results in a very short transmitter on-time. This architecture allows the sensors to consume very low energy resulting in long battery lives. The sensors also have onboard memory allowing them to function as a data logger.
Features:
- Onboard calibration table provides linear output
- Integrated 100mw, 900 MHz SSFH radio for long range
- Programmable log rates from 2 to 60 minutes
- Stores up to 3,072 data and/or event records
- User selectable Alarm limits
- LED indicates the following conditions: Alarm, Alarm Acknowledgement and Return to Normal
- Alarm Acknowledgement can be reset via a return radio transmission
- Return to Normal is used to allow the user to determine the exact duration of the alarm
- Configurable Alarm Utility with Audible and Visual Alarm Indication
- Transmission and alarm acknowledgement
- Wireless configuration
- Programmable transmit rates (4 minutes to 12 hours)
- Programmable log rates (2 minutes to 1 hour)
- Range Indoor: Up to 1300’
- Sensor input sampled every 15 seconds
Specifications:
- RTD power on Duty Cycle: 2%
- Measuring Current: 300 micro Amp @ 2% duty cycle
- Resolution: 0.1°C
- RTD Probe Operating Temperature: -50 to +125°C
- Sensor Accuracy +60 to -125°C: +/- 1.0°C
- Sensor Accuracy +30 to +60 and -50C to 0C: +/- 0.7°C
- Sensor Accuracy 0 to +30C: +/- 0.5°C
- Sensor Storage Temperature: -40 to +60°C
- Battery: 1.5 Vdc Lithium L91 x 4
- Battery Life: Up to 3 years
- Can be used with RF0515 Repeater for extended range
NOTE: Supplied with a 4" Stainless Steel RTD probe with 6' cable (pictured)