Wireless Professional Gen 2 Carbon Dioxide (CO2) Sensors


  • Part #:RFP-0285-2
  • Weight:4.5oz
  • Dimensions:4.375 x 2.470 x 1.12
  • Units:Inches
  • Availability: Inquire

Price: $378.00
Price: $510.30


Description

The mems based wireless carbon dioxide sensor allows you to monitor the level of carbon dioxide (CO2) gas in the surrounding air. Scigiene Wireless Professional Gen 2 Carbon Dioxide (CO2) Sensor have a small footprint and low cost but boast industry leading, premium performance specifications and are the longest lifetime sensors in the industry and the only CO2 sensor on the market powered by a coin cell battery (battery lasts over a year at 1 hour heartbeat). User customization allows you to set the frequency of readings and the ability to set customized alerts via SMS text or email when the sensor detects CO2 levels outside of the user's defined safe levels.

Features:

  • Wireless range of 1,200+ feet through 12+ walls
  • Frequency-Hopping Spread Spectrum (FHSS)
  • Improved interference immunity - Improved power management for longer battery life
  • Encrypt-RF® Security (Diffie-Hellman Key Exchange + AES-128 CBC for sensor data messages)
  • All Gen 2 sensors now have up to 3200 readings: 10-minute heartbeats = 22 days, 2-hour heartbeats = 266 days
  • Over-the-air updates (future proof)

Specifications:

Supply voltage

2.0-3.8 VDC (3.0-3.8 VDC using power supply)

Current Consumption

0.2 µA (sleep mode)

0.7 µA (RTC sleep)

570 µA (MCU Idle)

2.5 mA (MCU Active)

5.5 mA (Radio RX Mode)

22.6 mA (Radio TX Mode)

Operating temperature range (sensor and CO2 element)

0°C to 50°C (32°F to 122°F) standard range

 

Operating humidity range (sensor and CO2 element)

0% to 95% (non-condensing)

Recommended storage temperature (no batteries)

-30°C to +70°C (-22°F to +158°F)

Sensing method

Non-dispersive infrared (NDIR) absorption

Gold-plated optics

Solid-state source and detector

Sample method

Diffusion

Measurement range

0 to 10000 ppm CO2

Accuracy

+/-( (45 ppm + 3%) * (1 + .0013 * P)) 

Calibration

Autocalibration 

Non linearity

< 1% of FS

Pressure dependence

0.13% of reading per mm Hg in normal atmospheric conditions

Response time

3 minutes

Certifications

900 MHz product; FCC ID: ZTL-G2SC1 and IC:9794A-G2SC1.

Please call your Scigiene representative for help in designing your wireless system in order to make sure all range and other technical specifications are fully met.

Technical Information

Required Accessories

Scigiene 21 CFR Part 11 Premiere is a secure cloud based monitoring and notification software dedica (more...)

Part #: RFR-0539-2
$519.75 (USD)
$701.66 (CAD)
The Scigiene GEN 2 4G Industrial Cellular Gateway allows you to control GEN 2 Wireless Sensor settin (more...)

Part #: RFR-0538-2
$404.25 (USD)
$545.74 (CAD)
The Scigiene GEN 2 4G Cellular Gateway allows you to control GEN 2 Wireless Sensor settings without (more...)

Part #: RFR-0526-2
$231.00 (USD)
$311.85 (CAD)
The Gen 2 Ethernet Gateway allows your RF Wireless Sensors to communicate with the Scigiene Premier (more...)

Part #: RFR-0529-2
$288.75 (USD)
$389.81 (CAD)
The GEN2 Modbus Gateway acts as a data concentrator for Scigiene’s GEN 2 long range wireless sensor (more...)